亚洲av无码一区二区三区在线高,欧美人与性动交Α欧美精品,欧美激情精品久久,亚洲午夜无码毛片av久久久久久

熱門(mén)搜索:低溫冰箱、生化培養(yǎng)箱、鼓風(fēng)干燥箱、恒溫恒濕培養(yǎng)箱、箱式電阻爐、電子天平、滅菌器等等實(shí)驗(yàn)設(shè)備
欄目導(dǎo)航

avigation

網(wǎng)站首頁(yè)>技術(shù)文章 > NDJ-5S-8S-9SO Digital Viscometer perating Instruction

NDJ-5S-8S-9SO Digital Viscometer perating Instruction

Operating Instruction

-NDJ-5S-8S-9S

Digital Viscometer

CONTENT

1. Operating principle and Using …………………1

2. Main Technical Data ……………………………1

3. Application Environment ………………………2

4. Structher and Installation ………………………2

5. Using……………………………………………3

6. Panel operation…………………………………3

7. Precaution………………………………………6

8. Packing list………………………………………8  
1Operating priciple and Using

   NDJ Series Digital Viscometer (NDJ-5S-8S-9S) is a smart instrument,depending on the technology of A 16-bit high-performance single chip,which has totally changed the previous rotor method. Stepper motor runs accurately and stably as the setting speed procedure. And then torque sensor will drive rotor rotation in a constant speed. When rotor suffers viscous resistance in the tested liquid, the viscosity of the tested liquid will show on the screen by testing and handling.

   Compared with the same instrument, this one has the merit of high measure accuracy, stabled in rotate speed, wide working voltage (available in 110V, 220V, 50Hz, 60Hz) and so on. It is very convenient for users to choose the rotor and rotor velocity of the testing liquid correctly, because of the function of measured value occupies full range.

    It can widely used to measure Dynamic Viscosity of solvent based adhesive, latex, biochemicals, oil paint, coating, cosmetic, printing ink, pulp,starch, food and so on.

2: Main Technical Data

1、NDJ-1SNDJ-5S

    Measurement range10mPa.S-100,000mPa.S(1mPa.S=1cp)

    Measurement errorNDJ-1S:±5%,NDJ-5S:±3%,

    Rotor types1#, 2#, 3# and 4# rotors(0# is optional accessory)

    Rotor velocity6、1230、60RPM/min

    Dimensions95*130*155 (not including classis)

    Net weight2kg  (not including classis)

2、NDJ-8S

    Measurement range10mPa.S-2,000,000mPa.S(1mPa.S=1cp)

    Measurement error:±3%

    Rotor types1#, 2#, 3# and 4# rotors(0# is optional accessory)

    Rotor velocity0.3、 0.61.5、3、612、3060RPM/min

    Dimensions95*130*155 (not including classis)

    Net weight2kg(not including classis)

3、NDJ-9S

    Measurement range10mPa.S-6,000,000mPa.S(1mPa.S=1cp)

    Measurement error:±3%

    Rotor types1#, 2#, 3# and 4# rotors(0# is optional accessory)

    Measured Value will become larger as the rotor number grows

    Rotor velocity0.1、0.3、0.6、1.5、3、6、12、3060RPM/min

    Dimensions95*130*155(not including classis)

    Net weight2kg(not including classis)

3:Application Environment:

Environment temperature 5- 35

    Relative humidity:≤80%

    Power supply 220V,50Hz;110V,60Hz

    There is no electro magnetic interference, severe virbation, corrosive gas around the instrument.


4:Structure and Installation

 1Structure

            As follows:

1level babble  2screen  3cover  4rotor protection frame

5the host base 6operation key 7rotor connector  

8rotor9the host base horizontal knob

2Installation

1. You need to take classis, stanchion, lifting block and hand shank from the packing box. Screwing stanchions into the hole on the base. (The rack side must face to you), tiding the hand shank of viscomeler to the lifting block.

2. Adjust the tightness of up and down by swirling the up-down button, then putT"block behind instrument into hand shank and screw them.

3. Adjust three level screws of the base to make the level bubble in the center of  black circle.

4. Take off universal-joint protective cap under the main case.

5. Connect power

5:Operation

1. Prepare the liquid to be measured and put it into a round fat container or flat container with the diameter not smaller than 60mm, you need to control temperature of measured liquid correctly (It will influence viscometer directly)

2. Please reverse screw protection bracket beneath the main case.

3. Please choose the right rotor and contrarotate in the universal joint ofthe instrument

4. Adjust the lifting screw and put the rotor into the measured liquid till the level mark on the rotor reach the liquid surface.

5. Adjust the instrument in level again.

6. During the testing, the environment temperature must be stable to make sure the show value is accurate.


6Panel operation

This instrument has eight keys: choosing rotor, setting rotor speed,,,

stat, stop,search, confirm. Turn on the power switch, enter standby state.

7Precaution

1. Because Tem is viscosity's function, so the Tem measurement error should  be  controlled in ±0.1℃  when  the  instrument works in the normal temperaure.

2. Please look out the aperture angle percentage, its value should be kept 20%-85% . When the valuce  is  too  high  or  too  low,  please change rotor or velocity. Otherwise, it will affect measurement accuracy.

3. The instrument should be used in the required range of voltage  and frequency, Othervise, it will afect measurement accuracy.

4. You should operate carefully when you put on or take off rotor.  You just need to pull up the universal joints a little bit, forbidding  pushing too  hard.

5. Afer using, universal joints and rotor should be kept clean.

6. You need to drop the instrument slowly. You'd better hold it to  avoid damage the  axis because of vibration.

7. Universal joints should be put on protctive nap when the  instrument is moved or  transported.

8. Turbid liquid, emulsion, high polymer material and some high viscosity liquid, which all  belong non-Newtonian liquid. Their apparent viscosity and velocity are not linear relationship. So,  when  you  use  different rotors, velocity, and measurement time, their viscosity measurement may be different, not the mistake of instrument testing.

9. You will get accurate result when you do as fllowing:

    A. Control the tested liquid temperature accurately.

    B. Kcep environment temperature balance.

    C. Rotor and the tested liquid must have the same temperature.

    D. The bigger rotor, higher speed usually is used for low viscometer

     liquid, and high viscometer usually uscs smaller, lower speed.

    E. Keep the surface of rotor clean.

Attach: the installation step of  0#

1): The using of 0# must be followed as instument operations. You will    know the specification by picture.

2): Pouring 25-30mL tested liquid into sleeve which has bottom.

3): Put on the out sleeve from the bottom up and fix it. (Sleeve P9  has V-groove)

4): The same way to use sleeve. When the  out  sleeve  and  rotor  are immersed in liquid, taking the out sleeve  having  V-groove  inside as the liquid line of  0#.

Packing  list

No.           Name                               Number

1      NDJ series Digital Rotary Viscometer                 1

2      Stanchion and lifting block                          1

3     Protection bracket                               1

4      Rotors, 1#, 2#, 3# and 4#                          1

5     Power line                                      1

6      Operation manual                                1

7     Quality certificate                                1

8      Guarantee card                                  1

8Attach

Installation step:

Screw 0# rotor (5); fixed tap of sleeve (2); Screw in screw (1) and fix it;

Put on size zero sleeve (4); Screw in screw (1) and fix it.



TEL:13798128437

掃碼添加微信
久久99热狠狠色精品一区| 无码人妻久久一区二区三区免费丨| 久久久无码精品亚洲日韩蜜臀浪潮| 无限看片的视频高清在线| 少妇做爰免费视频网站| 国产精品宾馆在线精品酒店| 久久婷五月丁香色啪网| 野花日本免费完整版高清版 | 青青河边草免费视频观看| 久久久久久亚洲精品中文字幕 | 色欲人妻AAAAAA无码| 黑人巨茎大战欧美白妇| 无码少妇一区二区三区浪潮av| 亚洲欧美成人综合久久久| 亚洲欧美日韩综合一区| 久久精品国产亚洲77777| 国产chinese白嫩小受gv| 播放灌醉水嫩大学生国内精品| 久久婷婷五月综合97色一本一本| 亚洲av天堂丁香五色天 | 久久久久亚洲av无码专区首| 久久久久久亚洲精品中文字幕| 97人妻精品一区二区三区| 免费看少妇高潮成人片| 日韩AV无码精品一二三区 | 亚洲成a人片在线观看中文| 亚洲av久久无码精品九九| 国产无遮挡a片又黄又爽| 美女视频网站永久免费观看 | 国产线路3国产线路2| 亚洲国产成人片在线观看无码| 丰满少妇猛烈进入a片88| 一本加勒比hezyo无码专区| 亚洲综合久久精品无码色欲| 四川少妇被弄到高潮| 欧美VIVO2018| 女性自慰网站免费观看| 国产色视频一区二区三区| 野外做受又硬又粗又大视频√| 亚洲日日做日日谢日日鲁| 十大禁用的免费游戏手游中文|